4 | Description
INFICON
34 / 140
UL3000-Fab-Operating-instructions-iina73en1-08-(1910)
Local control
Unlocked
Clock style
Analog
Display background at standby
Off
Raise lower leak rate limit
0
Minimum volume
0
HYDRO•S (only UL3000 Fab PLUS)
Deactivated
Set maintenance interval
Sniffer tip filter
1000 h
Air filter
2500 h
Set user
Name
Supervisor
PIN
1111
Language
English
Auto Leak Test
Min. test time
5 s
Max. test time
10 s
Part numbering
ON
Part number
1
Warning limit for consecutive failed
tests
5
Diagram type
Line graph
Line graph
Scaling
Logarithmic
Decades
4
Autoscale
On
Lower diagram limit
1 × 10
-12
Time axis
30 s
Bar graph
Scaling
Logarithmic
Decades
4
Autoscale
On
Lower diagram limit
1 × 10
-12
Circle graph
Decades
5
Autoscale
On
Lower diagram limit
1 × 10
-12
Data recording