6 | Operation
INFICON
76 / 140
UL3000-Fab-Operating-instructions-iina73en1-08-(1910)
1
> Setup > Data recording
2
To record immediately after starting a measurement, activate the option
"Enabled".
Note: Data recording occurs during the measurements, i.e. while the
measurement screen is displayed. See also "Display measurement cycles
76]".
ð
This option is deactivated as standard. To start or stop a recording without
activating this option press
in the measurement display.
3
Under “Export format” select between "CSV en", "CSV de" and "JSON".
ð
Standard is "CSV en".
For "CSV en", the column separator is a comma.
For "CSV de", the column separator is a semicolon.
"JSON" is a language-independent data format.
ð
Export files have a default name with timestamp and extension (e.g.
icmeas20160720-082829.csv).
4
To export measurement data in addition to the measurement data, such as the
serial number of the device, activate the "Export metadata" option.
ð
Note: If other data is exported in addition to measurement data, this can
increase the effort required for data post-processing.
5
Save
.
6.6.2 Display measurement cycles
The measuring cycle is the time between pressing the START button and the STOP
button. You can view the associated information and optionally edit 2 free text fields.
1
> Measurement data > Measurement cycles
ð
Stored measurement cycles with ID, start and end time are displayed.
2
To view the details of a measurement cycle, select the desired entry by touching
it and press
.
ð
All information related to the measuring cycle is displayed.
3
To export the displayed cycle, connect a USB flash drive (FAT32 formatted) to
the leak detector and press
.
4
To edit text in the "Free text 1" and "Free text 2" fields, press
.
ð
Alternatively, you can edit "Free text 1" and "Free text 2" from the
"Measurement cycles" window. Select an entry by tapping on it and press
.
6.6.3 Export measurement data via network
In addition to access to the graphical user interface, you also have the option of
retrieving specific measurement data from your leak detector.
Summary of Contents for ULTRATEST UL3000 Fab
Page 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Page 133: ...INFICON Appendix 11 UL3000 Fab Operating instructions iina73en1 08 1910 133 140 ...
Page 134: ...11 Appendix INFICON 134 140 UL3000 Fab Operating instructions iina73en1 08 1910 ...
Page 138: ...Index INFICON 138 140 UL3000 Fab Operating instructions iina73en1 08 1910 ...
Page 139: ......
Page 140: ......