5.4 Connecting to the power supply system
WARNING
Danger from electric shock
Improperly grounded or fused products may be dangerous to life in case of a fault.
The use of the device is not permitted without a connected protective conductor.
► Only use the included 3-wire power cable.
► Only use the hospital grade power cable when using the device in a region with a
power supply of 100 to 120V.
► Always replace a defective power cord with a hospital grade power cord when using
the device in a region with a power supply of 100 to 120V.
► Make sure that the power supply plug is always accessible.
5.5 Checking the operation of the device
NOTICE
Damage to the turbo molecular pump due to jerking movements
Jerking movements can damage the running turbo molecular pump.
► Avoid any jerking movements or vibrations to the device during operation and for up
to 2 minutes after switching off.
Do not switch on the device when the ambient temperature is less than 10 °C.
ü
A DN 25 KF blank flange is available (if not already installed on the inlet flange).
ü
A helium calibration leak is available (optional).
1
Unpack the device, take a look at the supply and inspect for any visual damage.
2
Check if the inlet is blind-flanged. If this is not so, flange a blank flange with an
O-ring gasket on the inlet on the upper side of the device.
3
Connect the device to the power supply.
4
Switch on the leak detector via the power supply switch.
ð
After switching on pieces of status information are shown on the touchscreen
about the speed of the turbo molecular pump, the foreline pressure, the
emission and the active cathode. The start process takes about 3 minutes
and is completed with a short acoustic signal. Now, the device is in the
"Standby" mode (ready to operate).
5
Press the STARTbutton.
Summary of Contents for ULTRATEST UL3000 Fab
Page 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Page 133: ...INFICON Appendix 11 UL3000 Fab Operating instructions iina73en1 08 1910 133 140 ...
Page 134: ...11 Appendix INFICON 134 140 UL3000 Fab Operating instructions iina73en1 08 1910 ...
Page 138: ...Index INFICON 138 140 UL3000 Fab Operating instructions iina73en1 08 1910 ...
Page 139: ......
Page 140: ......