INFICON
Operation | 6
UL3000-Fab-Operating-instructions-iina73en1-08-(1910)
85 / 140
V10
V9
V8
V7
V5
V4c
V4b
V4a
V3
V2b
V2a
V1
p4
p2
p1
MS
TL
1
2
3
4
5
7
8
6
9
10
Fig. 13: The most important components of the vacuum diagram in UL3000 Fab devices
1 Mass spectrometer
6 Sniffer line
2 Internal calibration leak
7 Pressure measuring point for flow
monitoring
3 Inlet flange
8 Fore pump
4 Inlet pressure gauge
9 Fore-vacuum pressure gauge
5 V1 … V10: Solenoid valves
for controlling gas flows
10 Turbo molecular pump
6.12 Resetting to factory settings
You can restore the device to factory settings.
Loss of settings and measurement data
After resetting to factory settings only the manufacturer factory settings are in the
memory of the device. The software of the device is not reset when restoring the
factory settings.
ü
Supervisor - rights
1
> Setup > General > Reset
2
Make your choice:
ð
If you want to reset all settings, press the "Reset settings" button.
ð
If you want to delete all recorded measurement data and user accounts,
press the "Delete data" button.