6 | Operation
INFICON
70 / 140
UL3000-Fab-Operating-instructions-iina73en1-08-(1910)
After activating "Background suppression" the internal background and selectably also
the background in the inlet area is deducted from the leak rate. This happens after
pressing the START button automatically.
ü
Supervisor - rights
1
> Setup > Measurement > ZERO and filter
2
Make your selection in the field “Background suppression“ between “Internal
only”, “Inlet area” and “Switched off“.
ð
Internal only
Factory setting. With every measurement the internal measurement signal
component of the device is deducted by pressing the START button. The
accuracy of the measurement results are no longer impaired by the internal
background.
At the time of pressing the button START:
In order to determine a new internal background, the leak detector must be in
standby mode for at least 20 seconds and the flushing or gas ballast valve
must not have been open during this time. If you want to make sure that a
current internal ground is used in the following measuring operation, do not
press START, until the background display in standby has changed from low
opacity (grayed out) to normal opacity. To make the background display
visible in standby mode, activate the corresponding function, see “Change
53]”.
ð
Inlet area
Additionally to the deduction of the internal background, pressing START with
every measurement also deducts the background of the inlet area.
IF you would like to switch this on, make sure that the background of the inlet
area is determined, see "Determine the background of the inlet area [
71]".
We recommend that the calculation of the inlet area background is
occasionally repeated, because the device pumps down any background
present in the inlet area after a certain time.
ð
Switched off
No background suppression.
3
Save
.
High background value
If the actual background is higher than 1×10
-8
mbar l/s, it can no longer be
automatically deducted. The high background must be pumped down in this case.
See also
2
Set and use the function ZERO [
2
Summary of Contents for ULTRATEST UL3000 Fab
Page 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Page 133: ...INFICON Appendix 11 UL3000 Fab Operating instructions iina73en1 08 1910 133 140 ...
Page 134: ...11 Appendix INFICON 134 140 UL3000 Fab Operating instructions iina73en1 08 1910 ...
Page 138: ...Index INFICON 138 140 UL3000 Fab Operating instructions iina73en1 08 1910 ...
Page 139: ......
Page 140: ......