INFICON
Operation | 6
UL3000-Fab-Operating-instructions-iina73en1-08-(1910)
71 / 140
6.3.13.1 Determine the background of the inlet area
This function determines the value of the internal Helium background.
If "Zero and Filter" are shown in the window under background suppression with value
"Inlet area" selected, this value is then deducted from the measured signal
whenSTART is pressed. See also "Change background suppression [
69]".
ü
Inlet blank flange
ü
Operation mode Vacuum Mode
ü
The device is in the state "Standby" or "Measure".
1
Select the cross-hair icon
.
2
Select the calibration "Inlet background”.
3
To start the automatic determination of the intake background, press
.
6.3.14 Change SL3000 settings for sound / light (UL3000 Fab
PLUS only)
You can set up how you are to be additionally informed on the display of the
touchscreen concerning the measurements. The settings take effect with Setpoint 1.
ü
Operator or Supervisor rights
1
> Setup > Measurement > Sniff > SL3000 sniffer line
2
Select "Sniffer line light alarm configuration".
ð
As choices there are "Lighter", Off" or "Flash".
3
Select "Sniffer line light brightness".
ð
You have a choice of OFF / 1 - 6.
4
Switch-on the "Sniffer line beep" sniffer audio.
ð
As choices there are "Off" or "Beep when setpoint 1" available.
5
Save
.
6.3.15 Change the sniffer line pressure monitoring
In the operation mode Sniffer (SL200 or SL3000) automatic pressure monitoring is
activated. The preset pressure limits define a maximum and a minimum inlet pressure.
If the pressure does not lie within these limits then generally the caliper is either
blocked or broken. Fault messages are issued according to the following rules:
Inlet pressure > Upper limit: Caliper defective.
Inlet pressure < Lower limit: Gas flow is too low through the caliper (blocked caliper).
ü
Supervisor - rights
1
> Setup > Measurement > Sniff > Capillary surveillance
Summary of Contents for ULTRATEST UL3000 Fab
Page 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Page 133: ...INFICON Appendix 11 UL3000 Fab Operating instructions iina73en1 08 1910 133 140 ...
Page 134: ...11 Appendix INFICON 134 140 UL3000 Fab Operating instructions iina73en1 08 1910 ...
Page 138: ...Index INFICON 138 140 UL3000 Fab Operating instructions iina73en1 08 1910 ...
Page 139: ......
Page 140: ......