Enter a value of your choice in hours for the "Air filter" or press the adjacent
"Default value" button (2500 hours).
3
Save
.
After performing a maintenance, you should reset the desired period until the next
maintenance.
See also
2
Change the filter mat inside the tool box [
2
Replacement of the filter mat on the bottom of the device [
2
Replacing the SL200 sniffer line filter [
2
Replacing the SL3000 sniffer line filter [
6.2.12 Activate or deactivate maintenance requests
Enables or disables warnings when the time for maintenance is exceeded. The time
until the next maintenance is displayed.
ü
Supervisor - rights
1
> Setup > Maintenance requests
ð
In the “Enable/disable maintenance requests” window, you can activate or
deactivate the maintenance requests for the following modules:
- Calibration leak
- TMP
- Filter sniffer tip
- backing pump
- Exhaust filter
- Air filter
2
Set up.
3
Save
.
6.2.13 Activating or deactivating calibration requests
If the option "calibration request” is not active (factory setting), you only get a
calibration request in the following cases:
• The operation mode “Auto Leak Test“ has been switched on and the testing time
of Auto Leak Test has been changed.
• A SL3000 sniffer line with a new serial number has been connected (UL3000 Fab
PLUS only).
• The operation mode has been changed via the operating unit and there was no
calibration in the new operation mode within the last 24 hours.
Summary of Contents for ULTRATEST UL3000 Fab
Page 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Page 133: ...INFICON Appendix 11 UL3000 Fab Operating instructions iina73en1 08 1910 133 140 ...
Page 134: ...11 Appendix INFICON 134 140 UL3000 Fab Operating instructions iina73en1 08 1910 ...
Page 138: ...Index INFICON 138 140 UL3000 Fab Operating instructions iina73en1 08 1910 ...
Page 139: ......
Page 140: ......