4.3.2.4 Meaning of the button LEDs
START button LED
STOP button LED
Meaning
Off
Flashing red
No connection to the control unit
Pulsing blue
Pulsing blue
Run up
Off
Green
Standby vented (Vent valve open)
Off
Blue green
Standby pumped out (Vent valve
close)
Pulsing green
Off
Pump down
Off
Flashing green
Internal calibration
Flashing green
Off
External calibration
Green
Off
Measuring
Yellow (warning, faulty
measurement possible)
Off
Measuring with warning not
acknowledged
Off
Red
Error
ZERO button LED
Off
No measurement operation
Off
ZERO blocked
Blue
ZERO off and HYDRO•S off
Green
ZERO ON
Yellow
ZERO off and HYDRO•S on
4.3.2.5 Meaning of the status LED
The status LED is located on the back of the instrument inside of the connector block
for the equipment and control signals, see also "Connections for accessories and
29]”.
Status LED
Meaning
Flashing red
No communication with operating unit
Flashing blue
Run up
Blue
Standby
Flashing green
Evacuate (pumping)
Flashing green
Calibrating
Flashing green
Software is started (immediately after the
device is switched on)
Flashing green
Boot loader is waiting for a software
update (after starting the software update
of the basic unit)
Summary of Contents for ULTRATEST UL3000 Fab
Page 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Page 133: ...INFICON Appendix 11 UL3000 Fab Operating instructions iina73en1 08 1910 133 140 ...
Page 134: ...11 Appendix INFICON 134 140 UL3000 Fab Operating instructions iina73en1 08 1910 ...
Page 138: ...Index INFICON 138 140 UL3000 Fab Operating instructions iina73en1 08 1910 ...
Page 139: ......
Page 140: ......