7 | Warning and error messages
INFICON
90 / 140
UL3000-Fab-Operating-instructions-iina73en1-08-(1910)
Type
Notification
Possible sources of
error
Remedy
W132
Sniffer line type (XL) not
allowed
A sniffer line type (XL)
cannot be used with
this device
• Replace the sniffer line with a sniffer line
type supported by the device
W151
No communication with
operating unit
A software update or a
parameter reset has
been executed
• Confirm the warning message
• Check that the message does not appear
when you restart the device
Internal connection
problem between the
basic unit and the
operating unit
• Contact customer service
W152
No communication with
control panel
Internal connection
problem between the
basic unit and the
control panel
• Contact customer service
W171
CU1000 not supported
A CU1000 cannot be
used with this device
• Disconnect the CU1000 from this device
W201
Voltage U24V_MSB too low
Malfunction of 24V
power supply unit
• Contact customer service
Short circuit or
overload in the 24V
supply
• Contact customer service
W202
Voltage U24V_MSB too high Malfunction of 24V
power supply unit
• Contact customer service
W203
24V valve supply voltage out
of range
Malfunction of a valve
• Contact customer service
Short circuit or
overload in the 24V
valve supply
• Contact customer service
W206
24V operating unit supply
voltage out of range
Malfunction of
operating unit
• Contact customer service
Short circuit or
overload in the 24V
operating unit supply
• Contact customer service
W207
12V internal supply voltage
out of range
Short circuit or
overload in the 12V
supply
• Contact customer service
Summary of Contents for ULTRATEST UL3000 Fab
Page 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Page 133: ...INFICON Appendix 11 UL3000 Fab Operating instructions iina73en1 08 1910 133 140 ...
Page 134: ...11 Appendix INFICON 134 140 UL3000 Fab Operating instructions iina73en1 08 1910 ...
Page 138: ...Index INFICON 138 140 UL3000 Fab Operating instructions iina73en1 08 1910 ...
Page 139: ......
Page 140: ......