6 | Operation
INFICON
56 / 140
UL3000-Fab-Operating-instructions-iina73en1-08-(1910)
2
If necessary change the units of the "Leak rate unit vacuum", the "Leak rate unit
sniff" and the "Pressure unit".
ð
The option “Equal to display “ is activated according to the factory settings so
that the units for the interface can be shown exactly as the units of the device.
3
In order to independently select the units of the interface deactivate the option
"Equal to display.
ð
Further fields to set the units of the interface are shown.
4
To order to set the selected units if required, so that they can also be used for
the interface, activate the option "Equal to display".
5
Save
.
6.2.9 Change audio settings
In addition to the visual display of the measurement results you can change the
settings to the volume, to the Notification beep and to the audio alarm.
WARNING
Damage to hearing due to loud audio
The alarm level of the device can exceed 85 dB(A).
► Set a volume up to a maximum of "5".
► Use suitable hearing protection at set volumes above "5".
ü
Operator or Supervisor rights
1
> Audio
2
In order to change the volume of the audio alarm and the notification sounds,
select the desired "Volume". Do not fall below the set value for the “Minimum
volume“ in the next field.
ð
Setting range: 0 … 15
ð
You can listen to the set volume using the "Test" button.
3
If necessary, change the minimum volume.
ð
The "Minimum volume" is the volume for the audible alarm which cannot be
undershot. If you select a value greater than 0, the lowest volume level is
shown on the measurement screen after pressing the volume icon. This
allows you to prevent a non-authorized person from setting a value that is
below the ambient noise level.
ð
Setting range: 0 … 15
4
If you want to define the cause of an alarm or the type of notification beep more
accurately, then make your selection with the “Audio alarm mode” between
“Leak rate proportional”, “PINPOINT”, “SETPOINT”, “TRIGGER” and “Off”.
Summary of Contents for ULTRATEST UL3000 Fab
Page 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Page 133: ...INFICON Appendix 11 UL3000 Fab Operating instructions iina73en1 08 1910 133 140 ...
Page 134: ...11 Appendix INFICON 134 140 UL3000 Fab Operating instructions iina73en1 08 1910 ...
Page 138: ...Index INFICON 138 140 UL3000 Fab Operating instructions iina73en1 08 1910 ...
Page 139: ......
Page 140: ......