INFICON
Accessories and interfaces | 10
UL3000-Fab-Operating-instructions-iina73en1-08-(1910)
121 / 140
Function
State:
Description
Cathode 2
closed:
Cathode 2 is active
open:
Cathode 1 is active
Measuring
closed:
Measuring (ZERO is possible, all setpoint
outputs switch depending on the leak rate.)
open:
Standby or emission disabled (ZERO is not
possible, all setpoint outputs will return
"Leak rate threshold value exceeded".)
Open
open:
always open
Sniff
closed:
Operation mode Sniffing active
open:
Operation mode Vacuum active
Setpoint 1
closed:
Measured leak rate exceeds the setpoint 1
open:
Measured leak rate undercuts the setpoint 1
Setpoint 2
closed:
Measured leak rate exceeds the setpoint 2
open:
Measured leak rate undercuts the setpoint 2
Setpoint 3
closed:
Measured leak rate exceeds the setpoint 3
open:
Measured leak rate undercuts the setpoint 3
Setpoint 4
closed:
Measured leak rate exceeds the setpoint 4
open:
Measured leak rate undercuts the setpoint 4
Purge
closed:
Purge valve open
open:
Purge valve closed
Vent
closed:
Internal vent valve open.
open:
Internal vent valve closed
Standby or evacuate
closed:
State standby or evacuate active
open:
State standby of evacuate not active, such
as when measuring
Status calibration
leak valve
closed:
Calibration leak valve is open
open:
Calibration leak valve closed
Warning
closed:
Warning
open:
No warning
ZERO active
closed:
ZERO switched on
open:
ZERO switched off
10.2.5 Setting up the I/O module protocol
To switch between ASCII and LD protocol.
ü
Supervisor - rights
1
> Setup > Accessories > I/O module > I/O module protocol
Summary of Contents for ULTRATEST UL3000 Fab
Page 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Page 133: ...INFICON Appendix 11 UL3000 Fab Operating instructions iina73en1 08 1910 133 140 ...
Page 134: ...11 Appendix INFICON 134 140 UL3000 Fab Operating instructions iina73en1 08 1910 ...
Page 138: ...Index INFICON 138 140 UL3000 Fab Operating instructions iina73en1 08 1910 ...
Page 139: ......
Page 140: ......