INFICON
Cleaning and maintenance | 8
UL3000-Fab-Operating-instructions-iina73en1-08-(1910)
107 / 140
8
If you have activated the display of maintenance requests, set the maintenance
interval to the desired new period. Also see "Activate or deactivate maintenance
59] " and "Set maintenance interval "Filter sniffer tip" or "Air
58]".
Replacing the sinter
filter of the sniffer
handle
The sinter filter is in the sniffer handle. To replace the filter inserts, proceed as follows:
1
Switch off the leak detector.
2
Unscrew the two cross-head screws that hold the sniffer probe in place.
3
Remove sinter filter including o-ring.
Fig. 18: Sinter filter in the sniffer handle
4
Check the filter for visible soiling.
5
Insert a new sinter filter with o-ring in the base of the filter tip.
6
Tighten the sniffer tip again.
7
Switch on the leak detector.
8
Close the sniffer probe with your finger. You should be able to feel the
underpressure. If not then there is a leak somewhere and you have to check
sniffer probe and handle.
9
If necessary, set the maintenance interval to the desired new period. See also
"Set maintenance interval "Filter sniffer tip" or "Air filter" [
8.6 Creating screenshots
You can save the current screen contents of the device to an image file. You can use
such a file for service communication, for example.
1
Create a directory named "Screenshots" on a USB stick (FAT 32 formatted).
2
To take a screenshot, connect the USB stick to one of the USB ports of the leak
detector, see also the back of the “Control unit [
ð
A screenshot is automatically taken and stored in the directory on the USB
stick. The date and time are also saved.
3
To take another screenshot, disconnect the USB stick from the USB port and
then reconnect it to the leak detector.
Summary of Contents for ULTRATEST UL3000 Fab
Page 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Page 133: ...INFICON Appendix 11 UL3000 Fab Operating instructions iina73en1 08 1910 133 140 ...
Page 134: ...11 Appendix INFICON 134 140 UL3000 Fab Operating instructions iina73en1 08 1910 ...
Page 138: ...Index INFICON 138 140 UL3000 Fab Operating instructions iina73en1 08 1910 ...
Page 139: ......
Page 140: ......