INFICON
Operation | 6
UL3000-Fab-Operating-instructions-iina73en1-08-(1910)
55 / 140
6.2.7 Result display of the Auto Leak Test
1
3
2
5
4
7
6
Supervisor
1 "Passed" or "Failed" as the result of the last test under "Part number". Test
failed if the leak rate exceeds the setpoint.
2 Actual test time, within set limits, see “Set Auto Leak Test [
3 Part number, assigned with "Part numbering". Starting from a starting number,
the part numbers are automatically incremented, see “Set Auto Leak Test
63]“.
4 Failed tests in sequence are incremented to warning limit. After a passed test,
the defect count is reset to 0, see “Set Auto Leak Test [
63]”.
5 Auto Leak Test pause: When the test chamber is closed, you can interrupt the
measurements by pressing this button and confirming in the following window.
The test chamber remains in the pumped-down state.
6 Measurement result
7 Operation mode “Auto Leak Test” active, see “Select operation mode [
61]”
6.2.8 Change units
For the vacuum operation mode you can select between "Torr·l/s", "atm·cc/s", "Pa·m
3
/
s", and "mbar·l/s".
After switching the operation mode "Sniffer" in addition to the modes mentioned above
you can also select between the units "oz/yr", "g/a" and "ppm".
You can also choose between the following pressure units: “mbar”, “Pa”, “atm” and
“Torr”.
ü
Operator or Supervisor rights
1
> Setup > General > Units
Summary of Contents for ULTRATEST UL3000 Fab
Page 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Page 133: ...INFICON Appendix 11 UL3000 Fab Operating instructions iina73en1 08 1910 133 140 ...
Page 134: ...11 Appendix INFICON 134 140 UL3000 Fab Operating instructions iina73en1 08 1910 ...
Page 138: ...Index INFICON 138 140 UL3000 Fab Operating instructions iina73en1 08 1910 ...
Page 139: ......
Page 140: ......