INFICON
Accessories and interfaces | 10
UL3000-Fab-Operating-instructions-iina73en1-08-(1910)
117 / 140
10.2 I/O module
10.2.1 Create a connection between the device and the I/O
module
ü
Supervisor - rights
1
Connect the INFICON I/O module with a data cable to the LD socket on the rear
of the device, see "Connections for accessories and control signals [
2
> Setup > Accessories > Device selection
3
Select the "IO module".
4
Save
.
10.2.2 Configuring analog outputs
ü
Supervisor - rights
1
> Setup > Accessories > I/O module > Analog outputs configuration
2
Set-up the "Analog output 1".
ð
You can choose from "Off", Pressure p1", Pressure p2", "Leak rate mantissa",
"Leak rate exponent", "Leak rate linear", "Leak rate logarithmic", "Leak rate
mantissa hyst.", and "Via interface".
3
Set-up the "Analog output 2".
ð
You can choose from "Off", Pressure p1", Pressure p2", "Leak rate mantissa",
"Leak rate exponent", "Leak rate linear", "Leak rate logarithmic", "Leak rate
mantissa hyst.", and "Via interface".
4
Set-up the "Analog output scaling".
ð
You can choose from "0.5 V / decade", "1 V / decade", "2 V / decade", "2.5
V / decade", "3 V / decade", "5 V / decade", or "10 V / decade".
5
Set-up the "Exponent upper limit".
6
Save
.
The settings "Analog output scaling" and "Upper limit exponent" affect both analog
outputs dependent on the selected function, see the table overview below.
Functions and assignment of analog outputs:
Off
The analog outputs are
switched off
(Output voltage = 0 V).
Summary of Contents for ULTRATEST UL3000 Fab
Page 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Page 133: ...INFICON Appendix 11 UL3000 Fab Operating instructions iina73en1 08 1910 133 140 ...
Page 134: ...11 Appendix INFICON 134 140 UL3000 Fab Operating instructions iina73en1 08 1910 ...
Page 138: ...Index INFICON 138 140 UL3000 Fab Operating instructions iina73en1 08 1910 ...
Page 139: ......
Page 140: ......