Description | 4
UL3000-Fab-Operating-instructions-iina73en1-08-(1910)
23 / 140
Function
Touchscreen
Control panel
Stop
STOP button
Vent
Press the STOP button
again and keep it pressed
for approx. 2 seconds.
(Prerequisite is the setting
“ventilation mode”
“manual”, see "Activate
62]”.)
4.3.2.3 ZERO button
• To hide the "Background signal", see also "Definition of terms [
8]”.
• Can be operated via the touchscreen of the control panel.
• Using the ZERO button you can switch ZERO on and off. For further details see
"Set and use the function ZERO [
Function
Touchscreen
Control panel
Background suppression ON
ZERO button
Background suppression OFF
Press the ZERO button again and
keep it pressed for approx. 2
seconds.
The actual measured leak rate is considered to be a background signal. Depending on
the selected variant of operation mode the background signal is hidden until the
display limit of the respective vacuum range is reached.
Display limits according to factory setting:
1×10
-1
mbar l/s in MASSIVE
1×10
-7
mbar l/s in GROSS
1×10
-10
mbar l/s in FINE
1×10
-12
mbar l/s in ULTRA
To switch off the function ZERO again, press the buttonZERO for about 2 seconds.
The function ZERO should be activated only if the leak rate signal is stable and no
leak is measured. See also "Set and use the function ZERO [
See also
2
Summary of Contents for ULTRATEST UL3000 Fab
Page 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Page 133: ...INFICON Appendix 11 UL3000 Fab Operating instructions iina73en1 08 1910 133 140 ...
Page 134: ...11 Appendix INFICON 134 140 UL3000 Fab Operating instructions iina73en1 08 1910 ...
Page 138: ...Index INFICON 138 140 UL3000 Fab Operating instructions iina73en1 08 1910 ...
Page 139: ......
Page 140: ......