override switch on the processor is set. Qualified service technicians,
refer to “ADJUSTMENTS,” in the
HP 8360 B-Series Swept Signal
Generator/HP 8360 L-Series Swept CW Generator Service Guide
for
more information.
Universal
Error
Messages
Error Messages From -499 To -400
These error messages indicate that the Output Queue Control of
the swept CW generator has detected a problem with the message
exchange protocol. This type of error sets the Query Error Bit (bit
2) in the Event Status Register. One of the following has occurred:
n
An attempt has been made to read data from the Output Queue
when no output is present or is pending.
n
Data in the Output Queue has been lost.
Events that generate Query Errors do not generate Command Errors,
Execution Errors, or Device-specific Errors.
-440, Query UNTERMINATED after indefinite res
-430, Query DEADLOCKED
-430, Query DEADLOCKED;Output Buffer Full
-420, Query UNTERMINATED
-420, Query UNTERMINATED;Nothing To Say
-410, Query INTERRUPTED
Error Messages From -399 To -300
These error messages indicate that some device operations did not
properly complete, possibly due to an abnormal hardware or firmware
condition. This type of error sets the Device-specific Error (bit 3)
in the Event Status Register. Events that generate Device-specific
Errors do not generate Command Errors, Execution Errors, or Query
Errors.
-350, Too many errors and also -32768
-330, Self-test failed
-330, Self-test failed;Power-On Tests
-313, Calibration memory
Error Messages From -299 To -200
These error messages indicate that an error has been detected by
the swept CW generator’s Execution Control Block. An error of
this type sets the Execution Error Bit (bit 4) in the Event Status
Register. One of the following events has occurred:
n
A data element following a header was evaluated by the swept CW
generator as outside of its legal input range or is inconsistent with
the swept CW generator’s capability.
n
A valid program message can not be properly executed due to
some instrument condition.
Error Messages