l
Causes the sweep in progress to abort and reset. If
is
ON it immediately restarts the sweep. The pending operation flag
(driving
and *OPC?) undergoes a transition once the
sweep is reset.
l
CALibration:PEAKing:AUTO
l
CALibration:PEAKing:AUTO?
Sets and queries the automatic peaking function. If AUTO is ON, then
a peak is done at regular intervals automatically. After
the
setting is OFF.
l
ion :
:
Peaks the SYTM.
l
CALibration:PMETer:DETector:INITiate?
Initiates the specified calibration. These calibrations require the use
of an external power measurement. Once initiated, the swept CW
generator sets up for the first point to be measured, and responds to
the query with the frequency at which the power is to be measured.
The parameters mean:
Initiates a calibration of the internal detector logger
breakpoints and offsets.
Initiates a calibration of an external detector’s logger
breakpoints and offsets.
l
CALibration:PMETer:DETector:NEXT?
The parameter is the measured power that is currently produced
by the swept CW generator. You must supply this parameter after
measuring the power using an external power meter. The query
response is issued after the swept CW generator processes the
supplied parameter and settles on the next point to be measured.
The query response is:
The frequency [in Hz] that is currently produced.
0
The calibration is complete.
< o
An error has occurred and the calibration is aborted.
l
CALibration:PMETer:FLATness:INITiate?
Initiates the specified calibration. These calibrations require the use
of an external power measurement. Once initiated, the swept CW
generator sets up for the first point to be measured, and responds to
the query with the frequency at which the power is to be measured.
The parameters mean :
USER
Initiates a calibration at all of the user flatness points.
Initiates a calibration of the external flatness. Depends on
value of
ion :
:
S-24 Operating and Programming Reference