Sap
Function Group
SWEEP
M e n u M a p
Description
This
lets you set a sweep time for frequency sweeps or power
sweeps. The sweep time range is 10 ms to
but the fastest sweep
time is constrained by the frequency span. The fastest possible sweep
can be determined automatically:
1. Press SWEEP
this reveals the sweep menu keys.
Select more
, to scroll to the next page of the sweep menu.
Select
SwpTime Auto, to set
the sweep
time to automatic.
The
swept CW generator calculates the fastest possible calibrated
sweep time for any sweep span.
Whenever you press [SWEEP), the active entry area displays
the current sweep time and whether the sweep time is coupled to
the frequency span (far right hand side displays:
AUTO).
If the word
AUTO is not displayed then the sweep time auto function is off.
Programming Codes
SCPI:
SWEep[:FREQ uency]:TIME <num>[time suffix] or
Analyzer:
ST <num>
See Also
Power Sweep
“Power Level and Sweep Time Operation” in Chapter 1
“Programming Typical Measurements” in Chapter 1
Auto
Function Group
SWEEP
M e n u M a p
Description
This
lets you set the swept CW generator’s sweep time to a
minimum value for a chosen span and meet all specifications. The
sweep time is limited by a 300
sweep rate. An asterisk next
to the key label indicates this feature is active.
S-68 Operating and Programming Reference