Clear Memory
Clear Memory
Function Group
Menu Map
Description
Programming Codes
See Also
SYSTEM
8
This
causes the swept CW generator to return to the factory
preset instrument state, after writing alternating ones and zeroes over
all state information, frequency lists, and save/recall registers. You
can select the number of times to clear memory. When you select
Clear Memory, the swept CW generator displays the following in the
active entry area:
OF TIMES TO CLEAR MEMORY: X
Enter the number of times the state information should be
overwritten. While the swept CW generator is working to overwrite
the state information, it flashes the count on the display.
This
causes the swept CW generator to recall the original
calibration data stored in permanent memory (EEROM) all list and
user ALC correction data will be lost.
SYSTem:SECurity:COUNt <n>
SYSTem:SECurity[:STATe] ON
SYSTem:SECurity[:STATe] OFF
The transition from on to off triggers the blanking. Sending the “off
message by itself will do nothing.
Analyzer:
SHKZOHZ
Security Menu
“Using the Security Features” in Chapter 1
Operating and Programming Reference C-3