Access User Flatness Correction Menu
9. Press POWER
Select
Menu.
Select Delete Menu Delete All . This step insures that the
flatness array is empty.
11. Press
Leave the delete menu and return to the previous
soft key menu.
12. Select Copy List This step copies the frequency list into the
correction table in sequential order.
Select Freq Follow.
This sets the swept CW generator to CW
frequency mode to facilitate taking correction information. As
you scroll through the correction cells, the swept CW generator
produces the corresponding CW frequency at 0
14. Select Enter Corr . This
allows correction value entry.
15. Press
ON
/
OFF
).
This step enables user flatness correction.
16. For 5
set the appropriate power sensor cal factor on the
power meter.
17. Use the swept CW generator rotary knob to adjust for a
measurement of 0.00
on the power meter. Notice that a
correction value is entered at 5
18. Use the up arrow key to increment to the next correction cell.
19. For 11
set the appropriate power sensor cal factor on the
power meter.
20. Use the swept CW generator rotary knob to adjust for a
measurement of 0.00
on the power meter.
21. Repeat this sequence of steps until all the frequency points have
a correction value entered.
Activate List Mode
22. Press SWEEP
Select Sweep Mode List .
23. The flatness correction array is ready to be applied to your setup.
Disconnect the power meter/sensor. The power produced at the
point where the power meter/sensor was disconnected is now
calibrated at the frequencies and power level specified above.
1-38 Getting Started Advanced