Scalar Analysis Measurement with User Flatness Corrections,
Example 4
The following example demonstrates how to set up a scalar analysis
measurement (using an HP 8757 Scalar Network Analyzer) of a 2 to
20
test device such as, an amplifier. User flatness correction
is used to compensate for power variations at the test port of a
directional bridge. Follow the instructions to set up the swept CW
generator, then configure the system as shown in Figure 1-21.
Note
The swept CW generator’s rear panel language and address switches
must be set to 7 and 31 (all l’s), to change the language or address
of the swept CW generator from the front panel. The programming
language must be set to Analyzer. Refer to menu map 8, System,
to find the location of
Programming Language Analyzer
(asterisk on active language).
RF OUT
POUER
c u
SCRLRR
GENERATOR
NETUORK RNRLYZER
I
J
‘PORT
Figure l-21. Scalar System Configuration
Getting Started Advanced 1-43