Det Cal
Function Group
Menu Map
Description
Programming Codes
See Also
ENTRY
NONE
This
lets you turn off (blank) the active entry area and
disable the ARROW keys, rotary knob, and entry keys. When
any function key (hard or soft) is pressed, the active entry area is
reactivated. The yellow LED, ENTRY ON, next
to (
ENTRY O N
/
OFF
)
indicates whether the entry area is active (LED on=active).
SCPI:
No specific code activates
(
ENTRY O N
/
OFF
)
Arrow Keys
“Entry Area” in Chapter
1
Ext Det Cal
Function Group
USER CAL
M e n u M a p
Description
This
enables the swept CW generator to act as a controller to
an HP 437B power meter. This
causes an immediate execute
on the interface bus and generates an HP-IB error if no power meter
is present on the interface bus or if the swept CW generator is unable
to address the power meter. Use external detector calibration to
characterize and compensate for an external negative diode detector
used in an external leveling configuration.
Programming Codes
CALibration:PMETer:DETector:INITiate?
CALibration:PMETer:DETector:NEXT? <num>[lvl suffix]
Analyzer:NONE
See Also
“Optimizing Swept CW Generator Performance” in Chapter
1
Operating and Programming Reference E-5