Sets the HP-IB address of the printer to use during some of the
calibration procedures when the swept CW generator assumes HP-IB
control. *RST and power on do not effect this command. The
default is 1. The default value is set at memory initialization only.
l
DIAGnostics:IORW
Performs a write to the I/O Device number specified in the first
and sets it to the value in the second
l
DIAGnostics:IORW?
Reads from the specified I/O device number and returns the response
data.
l
DIAGnostics:OUTPut:FAULt?
Returns a string of 16, and that are equivalent to the fault
display.
Bit 0 = PEAK
Bit 1 = TRACK
Bit 2 RAMP
Bit 3 = SPAN
Bit 4 =
Bit 5 = ADC
Bit 6 = EEROM
Bit 7 = PWRON
Bit 8 = CALCO
Bit 9 PLLZERO
Bit 10 = PLLWAIT
Bit 11 = FNXFER
Bit 12 = CAL YO
Bit 13 = CAL MAN
Bit 14 TMR CNFLCT
Bit 15 = SEARCH
DIAGnostics:RESult?
Returns the following information:
<result>, <test failure>, <manual entry point>
where, <result> is one
of the following:
0
Diagnosis successful.
1
Cannot diagnose; full
must be executed first.
2
No failures found-all selftests passed.
- 1
Cannot diagnose; diagnosis routine failed to isolate failure.
(software fault)
<test
failure> is
the
test number of the most relevant failure
(-999
if parameter is not used, as in <result> of 1).
entry point> is a string response that identifies the
paragraph number in the
HP 8360 B-Series Swept Signal Generator/
Operating and Programming Reference S-27