MHz Freq Std
auto
Function Group
SYSTEM
M e n u M a p
Description
This
sets the swept CW generator to choose its frequency
standard automatically. If an external standard
is
connected to the
10 MHz REF INPUT BNC, then it is chosen as the reference. If no
external standard is connected, the internal standard is chosen as the
reference. If the internal standard has been disconnected also, the
swept CW generator operates in a free run state. An asterisk next to
the key label indicates that this feature is active.
Programming Codes
SCPI:
ROSCillator[:SOURce]:AUTO
Analyzer:
NONE
See Also
Ref Osc Menu
MHz Freq Std
Extrnl
Function Group
SYSTEM
M e n u M a p
Description
This
tells the swept CW generator to accept an external
10 MHz signal as the frequency reference. The external signal must
be applied to the
10
MHz REF INPUT BNC connector located on
the rear panel. If no external signal is applied,
UNLOCK
and
EXT REF
appears on the message line of the display. An asterisk next to the
key label indicates that this feature is active.
Operating and Programming Reference T-l