List
Menu
LIST[:POWer]:CORR
ec
In the above three commands, the swept CW generator responds
with the number of points for the named parameter that are in the
list array.
If a particular list is shorter than another, an error is generated upon
execution. An exception is made for the case where the shorter list
is of length 1. In this case, the list of length 1 is treated as though
it were a list of equal length, with all values the same. At
all
lists for the current state are cleared and reset to a single value.
Analyzer:
NONE
Sweep Mode List
“Creating and Using a Frequency List” in Chapter 1
List Mode Pt Trig
Auto
Function Group
Menu Map
Description
Programming Codes
See Also
FREQUENCY
2
This
lets you set the swept CW generator to automatically
step through a frequency list, when the swept CW generator is in
sweep list mode.
SCPI:
LIST:TRIGger:SOURce
Analyzer:
NONE
List Menu, Pt Trig Menu, Sweep Mode List
“Creating and Using a Frequency List” in Chapter 1
L-8 Operating and Programming Reference