state. Programming a specific value for POWer :
ier : STATE sets
POWer:AMPLifier:STATE:AUTOto OFF.
is ON.
l
POWer:ATTenuation
l
POWer :
ion?
Sets and queries the output attenuation level. Note that when
setting the attenuator level to 10
the output power is
decreased by 10
Programming a specified attenuation sets
POWer :
ion : AUTO OFF.
l
POWer:ATTenuation:AUTO
POWer:ATTenuation:AUTO?
Sets and queries the state of the RF attenuator coupling switch.
Programming a specified attenuation sets
POWer:ATTenuation:AUTO OFF.
ON insures that the amplitude level of the ALC is kept within
optimal limits.
OFF the attenuator setting is set to the value of POW :ATT and left
there. The *RST value is ON.
l
POWer:CENTer
suffix]
l
POWer:CENTer?
Sets and queries the center power for power sweep. Default units
(and units for query response) are determined by the UNIT:POWer
command.
The coupling equations for power sweep are exactly analogous to
those for frequency sweep. Power sweep is allowed to be negative,
unlike frequency sweeps. See
: CENT for a description. *RST
value is 0
l
suffix]
l
POWer :
?
Sets and queries the output level. Default units and units
for the query response are determined by the UNIT:POWer
command. Maximum and minimum levels refer to the
leveling mode at the time the command is sent. For example,
POWer :
MIN; ALC :
MMHead has different effects
from
MMHead; POWer:LEVelMIN
After
the value is 0
l
POWer:MODE
l
POWer : MODE?
Sets and queries the setting of the power sweep mode switch. If in
the sweep mode then the output level is controlled by the start,
stop, center and span functions. If in the fixed power mode then the
output is controlled by the POW C: LEVEL] command. The *RST value
is
Operating and Programming Reference S-37