P
Peak RF Always
Function Group
POWER, USER CAL
M e n u M a p
Description
This
appears in two locations: the POWER
Tracking Menu
and the USER CAL
Tracking Menu.
The
operation is
the same in
both locations.
This
causes the swept CW generator, when in CW or
manual-sweep output mode, to align the output filter (SYTM) so
that its
is centered on the RF output. Peaking is used to
obtain both the maximum available power and spectral purity, and
the best pulse, FM envelopes, at a given frequency. This peaking
occurs each time the frequency is changed, or every seven minutes.
An asterisk next to the key label indicates this function is active.
Programming Codes
SCPI:
CALibration:PEAKing:AUTO
Analyzer:
function on, RPO function off.
See Also
Auto Track, Peak RF Once, Tracking Menu
“Optimizing Swept CW Generator Performance” in Chapter 1
Operating and Programming Reference P-l