Marker
Function Group
MARKER
M e n u M a p
Description
The
labeled Marker Ml through Marker M5 function
identically. The
turns the marker off/on. When an asterisk
appears next to the key label, it indicates that the marker is on, but
not necessarily active. A marker is only active when it is indicated in
the active entry area.
The active entry area displays the active marker and its frequency
value. Use the rotary knob, the ARROW keys, or the entry keys to
set the frequency. Markers are displayed normally as Z-axis intensity
dots, but can be changed to amplitude pulses (Amp1 Markers
When a marker is turned off, the frequency value of that marker
is retained in memory. If the marker is reactivated, the stored
frequency value is recalled for that marker.
The frequency value of Ml and of M2 can also be used to define
parameters in two other marker features:
Ml--M2
Sweep and
.
Programming Codes
MARKer[l][:FREQ uency] <num>
suffix] or
MARKer[l]:STATe
Analyzer: Ml function on, MO function off.
See Also
Markers, Ml--M2 Sweep ,
Menu ,
Start=Ml
“Marker Operation” in Chapter 1
“Programming Typical Measurements” in Chapter 1
M-4 Operating and Programming Reference