Enclosure Protection
Power Requirements
Line Voltage and Fuse
Selection
CAUTION
Position the instrument according to the enclosure protection
provided.
n
This instrument does not protect against the ingress of water.
n
This instrument protects against finger access to hazardous parts
within the enclosure.
The HP 8360 L-Series swept CW generators require a power source of
115 V
o
or 230 V
48 to 66 Hz, single-phase.
Power consumption is 400 VA maximum (30 VA in standby).
The swept CW generator is provided with a voltage selector (located
on the rear panel) to match the swept CW generator to the ac line
voltage available at the site of installation. Both the line selector
and fuse were selected at the factory to match the ac line voltage
expected to be found at the shipping destination. Verify that the
voltage selector has been set to the correct line voltage before
connecting power to the swept CW generator.
For continued protection against fire hazard replace line fuse only with
same type and rating. The use of other fuses or material is prohibited.
Refer to “Routine Maintenance” in Chapter 4 for information on
changing fuses.
Before switching on this product, make sure that the line voltage
selector switch is set to the voltage of the power supply and the
correct fuse is installed. Assure the supply voltage is in the specified
range.
Installation 3-3