The save/recall registers store and access a previously set instrument
state.
For example, set the swept CW generator to sweep from 3 to
15
at a -10
power level, with markers
1
and 2 set at 4.5 and
11.2
Press (START)
Press
Press
(
P O W E R L E V E L
)
Press (MARKER).
Select
Marker
Ml
Select
Marker M2
To save this instrument state in register 1, press
To verify
that the swept CW generator has saved this state:
Press [PRESET).
Press
Press (MARKER).
The active entry area displays:
RECALL REGISTER: 1 RECALLED
Notice the sweep end points, power level, and the asterisks next to
the marker 1 and 2 key labels.
You can save instrument states in registers 1 through 8. Register
0 saves the last instrument state before power is turned off. When
power is turned on, register 0 is automatically recalled.
1-16 Getting Started Basic