89
Analog Triggers, Pulse Triggers, Waveform Calculation Triggers (Level, Window)
Unit
Measure
Channel
Trigger
Alarm
Calculation
System
Level triggers
Level triggers activate when the waveform crosses the specified level (the trigger level).
You can set the direction in which the level is crossed (the slope).
>
>
[Unit n]
(n = 1, 2, . . .)
2
2
3
3
1
1
1
1
1
Under
[Type]
for
[Start]
or
[Stop]
, set the trigger type to
[Level]
.
A level trigger will be used.
2
Select the slope.
The trigger will activate when the level is crossed in the set direction.
↑
Activates the trigger when the level is crossed the specified level from below. (Rising)
↓
Activates the trigger will activate when the level is crossed the specified level from
above. (Falling)
When
[Condition]
is set to
[AND]
, the system will determine whether the waveform has exceeded the
specified level.
3
Under
[Level]
, set the trigger level.
The trigger will activate when the set level (temperature, voltage, etc.) is crossed.
When using the scaling function, post-scaling values are displayed.
(When a strain measurement module is used)
The instrument expresses strain in terms of micro epsilon (μ
ε
). You do not need to enter the SI
prefix micro (μ).
2
Trigger Function
Summary of Contents for LR8450
Page 68: ...64 Observing Waveforms Alarm...
Page 72: ...68 Observing Waveforms 3 Numerical calculation Numerical calculation results...
Page 131: ...128 Acquiring Data with a Computer PC...
Page 175: ...172 Controlling the System...
Page 233: ...230 Controlling the Instrument with Communication Commands...
Page 271: ...268 Plug in Module Specifications...