258
Plug-in Module Specifications
Measurement range, maximum resolution, measurable range, and measurement accuracy figures
Measurement
target
Type
Range
Maximum
resolution
Measurable range
Measurement
accuracy
Voltage
–
10 mV f.s.
500 nV
−10 mV to 10 mV
±10 µV
20 mV f.s.
1 μV
−20 mV to 20 mV
±20 µV
100 mV f.s.
5 μV
−100 mV to 100 mV
±50 µV
200 mV f.s.
10 μV
−200 mV to 200 mV
±100 µV
1 V f.s.
50 μV
−1 V to 1 V
±500 µV
2 V f.s.
100 μV
−2 V to 2 V
±1 mV
10 V f.s.
500 μV
−10 V to 10 V
±5 mV
20 V f.s.
1 mV
−20 V to 20 V
±10 mV
100 V f.s.
5 mV
−100 V to 100 V
±50 mV
1-5 V f.s.
500 μV
1 V to 5 V
±5 mV
Thermocouple
(Not including
accuracy of
reference
junction
compensation)
K
100°C f.s.
0.01°C
Greater than or equal to −100°C
and less than 0°C
±0.7°C
Greater than or equal to 0°C and
less than or equal to 100°C
±0.5°C
500°C f.s.
0.05°C
Greater than or equal to −200°C
and less than −100°C
±1.4°C
Greater than or equal to −100°C
and less than 0°C
±0.7°C
Greater than or equal to 0°C and
less than or equal to 500°C
±0.5°C
2000°C f.s. 0.1°C
Greater than or equal to −200°C
and less than −100°C
±1.4°C
Greater than or equal to −100°C
and less than 0°C
±0.7°C
Greater than or equal to 0°C and
less than 500°C
±0.5°C
Greater than or equal to 500°C
and less than or equal to 1350°C
±0.7°C
J
100°C f.s.
0.01°C
Greater than or equal to −100°C
and less than 0°C
±0.7°C
Greater than or equal to 0°C and
less than or equal to 100°C
±0.5°C
500°C f.s.
0.05°C
Greater than or equal to −200°C
and less than −100°C
±0.9°C
Greater than or equal to −100°C
and less than 0°C
±0.7°C
Greater than or equal to 0°C and
less than or equal to 500°C
±0.5°C
2000°C f.s. 0.1°C
Greater than or equal to −200°C
and less than −100°C
±0.9°C
Greater than or equal to −100°C
and less than 0°C
±0.7°C
Greater than or equal to 0°C and
less than or equal to 1200°C
±0.5°C
Summary of Contents for LR8450
Page 68: ...64 Observing Waveforms Alarm...
Page 72: ...68 Observing Waveforms 3 Numerical calculation Numerical calculation results...
Page 131: ...128 Acquiring Data with a Computer PC...
Page 175: ...172 Controlling the System...
Page 233: ...230 Controlling the Instrument with Communication Commands...
Page 271: ...268 Plug in Module Specifications...