166
Configuring Settings
8
Under
[
Power frequency filter]
, select the power supply frequency in the region where the
instrument is being used.
60 Hz
Uses the digital filter for 60 Hz regions.
50 Hz
Uses the digital filter for 50 Hz regions.
It is recommended to select the same frequency (50 Hz or 60 Hz) as the power supply frequency in the region
where the instrument is being used.
When using the U8550 to U8553 modules, power supply noise can be rejected by using the data refresh
interval setting.
See "Measurement module data refresh intervals" (p. 13).
Configure the low pass filter for Strain Units (U8554, LR8534) since the power supply frequency filter function
is not available for those modules.