302
Data Handling
11.13
Data Handling
In the following circumstances, calculated values and saved data are treated as shown in the table
below:
•
When the waveform significantly exceeds the range’s measurable range (+OVER, -OVER)
• When communications are temporarily interrupted (NO DATA)
• When the instrument detects a thermocouple wire break during temperature measurement (wire
break detection)
Input type
Input range
+OVER
−OVER
NO DATA
Wire break
detection
Voltage
1 mV
0.00163835
- 0.0016384
0.00163825
–
2 mV
0.0032767
- 0.0032768
0.0032765
–
5 mV
0.00819175
- 0.008192
0.00819125
–
10 mV
0.0163835
- 0.016384
0.0163825
–
20 mV
0.032767
- 0.032768
0.032765
–
50 mV
0.0819175
- 0.08192
0.0819125
–
100 mV
0.163835
- 0.16384
0.163825
–
200 mV
0.32767
- 0.32768
0.32765
–
1 V
1.63835
- 1.6384
1.63825
–
2 V
3.2767
- 3.2768
3.2765
–
10 V
16.3835
- 16.384
16.3825
–
20 V
32.767
- 32.768
32.765
–
100 V
163.835
- 163.84
163.825
–
1-5 V
16.3835
- 16.384
16.3825
–
Thermocouple
100°C
327.67
- 327.68
327.65
327.66
500°C
1638.35
- 1638.4
1638.25
1638.3
2000°C
3276.7
- 3276.8
3276.5
3276.6
Resistance
temperature
detector
100°C
327.67
- 327.68
327.65
–
500°C
1638.35
- 1638.4
1638.25
–
2000°C
3276.7
- 3276.8
3276.5
–
Humidity
100% RH
3276.7
- 3276.8
3276.5
–
Resistance
10
Ω
16.3835
- 16.384
16.3825
–
20
Ω
32.767
- 32.768
32.765
–
100
Ω
163.835
- 163.84
163.825
–
200
Ω
327.67
- 327.68
327.65
–
Strain
1000 μ
ε
1638.35
- 1638.4
1638.25
–
2000 μ
ε
3276.7
- 3276.8
3276.5
–
5000 μ
ε
8191.75
- 8192
8191.25
–
10000 μ
ε
16383.5
−16384
16382.5
–
20000 μ
ε
32767
−32768
32765
–
50000 μ
ε
81917.5
−81920
81912.5
–
100000 μ
ε
163835
−163840
163825
–
200000 μ
ε
327670
−327680
327650
–
Integration
1000 Mc
2147483647
–
–
–
Summary of Contents for LR8450
Page 68: ...64 Observing Waveforms Alarm...
Page 72: ...68 Observing Waveforms 3 Numerical calculation Numerical calculation results...
Page 131: ...128 Acquiring Data with a Computer PC...
Page 175: ...172 Controlling the System...
Page 233: ...230 Controlling the Instrument with Communication Commands...
Page 271: ...268 Plug in Module Specifications...