22
Configuring Input Channels
6
Under
[Burn out]
, select whether you wish to detect wire breaks.
OFF
Does not detect thermocouple wire breaks.
Values will vary when a thermocouple experiences a wire break.
ON
Detects thermocouple wire breaks when measuring temperature using thermocouples.
The numerical value display and cursor value will be indicated as
[BURNOUT]
when a
wire break occurs. Calculated values and saved data are treated as 327.66°C (100°C
f.s. range), 1638.3°C (500°C f.s. range), or 3276.6°C (2000°C f.s. range).
There are limits on the data refresh interval that can be set.
See “Thermocouple wire break detection” (p. 23).
7
Under
[RJC]
, select the type of reference junction compensation to use.
INT
Performs reference junction compensation inside the measurement module.
Use this setting when the thermocouple (or compensation lead wire) is connected
directly to the instrument.
The measurement accuracy is determined by adding the temperature measurement
accuracy to the reference junction compensation accuracy.
EXT
Does not perform reference junction compensation inside the measurement module.
Use this setting when connecting an external zero junction compensation device (0°C
ice water, etc.).
The measurement accuracy is defined by the temperature measurement accuracy
alone.
Measurable temperature range
The measurable temperature range depends on the type of thermocouple being used.
Thermocouple
Measurable temperature range
K
−200°C
to
1350°C
J
−200°C
to 1200°C
E
−200°C
to 1000°C
T
−200°C
to
400°C
N
−200°C
to 1300°C
R
0°C to
1700°C
S
0°C to
1700°C
B*
400°C
to
1800°C
C
0°C to 2000°C
*:
[B]
can be selected when using the 2000°C range. Temperatures from 0°C to 400°C will be displayed even
if
[B]
is selected, but accuracy will not be guaranteed.
Summary of Contents for LR8450
Page 68: ...64 Observing Waveforms Alarm...
Page 72: ...68 Observing Waveforms 3 Numerical calculation Numerical calculation results...
Page 131: ...128 Acquiring Data with a Computer PC...
Page 175: ...172 Controlling the System...
Page 233: ...230 Controlling the Instrument with Communication Commands...
Page 271: ...268 Plug in Module Specifications...