41
Configuring the Waveform Display
Unit
Measure
Channel
Trigger
Alarm
Calculation
System
2
Under
[Display horizontal axis]
, select the display format to use for time values (horizontal-
axis values) displayed on the screen.
Time
Displays the time elapsed since the start of measurement.
When using a trigger, the time elapsed since the trigger activation time will be
displayed.
Date
Displays the actual time (date and time) every 10 divisions.
Data points
Displays the number of data points since the start of measurement.
When using a trigger, the number of data points from the trigger activation time will be
displayed.
This setting also applies when displaying time values for waveform data saved in the CSV format.
3
Under
[Number display format]
, select the display format to use for measured values
(vertical-axis values).
Standard
Displays measured values using the same SI prefix as the range.
Example: 0.01234V (when using the 1 V range)
Decimal
Displays measured values as decimals.
Example: 0.012V (when the number of decimal places is set to 3)
Exponent
Displays measured values as exponents.
Example: 1.234E-02V (when the number of decimal places is set to 3)
Prefix
Displays measured values using an SI prefix.
Example: 12.345mV (when the number of decimal places is set to 3)
When setting the horizontal axis on the waveform screen
Under
[Horz]
, you can select the time per division. This setting can be changed while measurement
is in progress.
1
Settings and Operation
Summary of Contents for LR8450
Page 68: ...64 Observing Waveforms Alarm...
Page 72: ...68 Observing Waveforms 3 Numerical calculation Numerical calculation results...
Page 131: ...128 Acquiring Data with a Computer PC...
Page 175: ...172 Controlling the System...
Page 233: ...230 Controlling the Instrument with Communication Commands...
Page 271: ...268 Plug in Module Specifications...