20
Configuring Input Channels
5
(For the U8554 or LR8534 Strain Unit)
Under
[Filter]
, select the cutoff frequency.
Auto
,
120 Hz
,
60 Hz
,
30 Hz
,
15 Hz
,
8 Hz
,
4 Hz
See the table in “Measuring strain” (p. 27) for a list of cutoff frequencies when
[Auto]
is selected.
When measuring instrumentation devices
•
When measuring a 4-20 mA current, connect a 250
Ω
resistor between the positive and negative
input terminals.
See “Connecting voltage cables and thermocouples” in the Quick Start Manual.
• The
[1-5 V]
range is convenient when measuring output from 4-20 mA instrumentation devices.
• The
[1-5 V]
range is a range in which the lower and upper limits of the
[10 V]
range’s display
range are automatically set to 1 V and 5 V, respectively. If you wish to change the upper and
lower limit values, use the
[10 V]
range.
•
The Strain Unit does not support 4-20 mA current measurement.
You can use the scaling function to convert measured voltage values to the desired values.