249
Plug-in Module Specifications
U8551 Universal Unit
1.
General specifications
Compatible logger
LR8450/LR8450-01 Memory HiLogger
Operating
environment
Indoors, Pollution Degree 2, altitude up to 2000 m (6562 ft.)
Operating temperature
and humidity range
−10°C to 50°C (14°F to 122°F), 80% RH or less (non-condensing)
Storage temperature
and humidity range
−20°C to 60°C (−4°F to 140°F), 80% RH or less (non-condensing)
Standards
Safety: EN 61010
EMC: EN 61326, Class A
Standard compliance
Thermocouples: JIS C1602:2015, IEC 60584-1:2013
Resistance temperature detector (Pt100, Pt1000): JIS C1604:2013, IEC 60751:2008
Resistance temperature detector (JPt100): JIS C1604:1989
Vibration resistance
JIS D 1601:1995 5.3 (1)
Class 1: Passenger vehicles; conditions: Class A equivalent
(4 h along X-axis and 2 h along Y- and Z-axes at a vibration acceleration of 45 m/s
2
[4.6 G])
Dimensions
Approx. 134W × 70H × 63D mm (5.28″W × 2.76″H × 2.48″D) (including cover)
Mass
Approx. 318 g (11.2 oz.)
Product warranty
period
3 years
Accessories
Instruction Manual
Installation screws × 2
Options
Z2000 Humidity Sensor
2.
Input, output, and measurement specifications
-1.
Basic specifications
Number of input
channels
15 (Set voltage, thermocouple, humidity, resistance temperature detector, or resistor for
each channel.)
Input terminals
Push-button type terminal block (4 terminals per channel), outfitted with terminal block
cover
Measurement target
Voltage
Thermocouples (K, J, E, T, N, R, S, B, C)
Humidity (using the Z2000 Humidity Sensor)
Resistance temperature detector (Pt100, Jpt100, Pt1000)
(Connection: 3-wire/4-wire)
(Measurement current: 1 mA ±5% [Pt100 and Jpt100 measurement];
0.1 mA ±5% [Pt1000 measurement])
(During Pt1000 measurement, setting available when the data refresh interval is
greater than 100 ms.)
Resistor (connection: 4-wire; measurement current: 1 mA ±5%)
Input type
Scanning by semiconductor relays, floating unbalanced input
All channels isolated (The SoL terminals used to measure resistance temperature
detectors and resistors are not isolated as they are shorted for all channels internally.)
A/D resolution
16 bits
Maximum input
voltage
±100 V DC
Maximum channel-to-
channel voltage
300 V DC (The SoL terminals used to measure resistance temperature detectors and
resistors are not isolated as they are shorted for all channels internally.)
10
Specifications
Summary of Contents for LR8450
Page 68: ...64 Observing Waveforms Alarm...
Page 72: ...68 Observing Waveforms 3 Numerical calculation Numerical calculation results...
Page 131: ...128 Acquiring Data with a Computer PC...
Page 175: ...172 Controlling the System...
Page 233: ...230 Controlling the Instrument with Communication Commands...
Page 271: ...268 Plug in Module Specifications...