Configuring Input Channels
Unit
Measure
Channel
Trigger
Alarm
Calculation
System
Measuring strain
This section describes how to configure settings on the individual settings screen when measuring
strain or vibration with a strain gage or strain gage-type converter.
You can use
[Input]
on the settings list screen to configure the settings. (See p. 50.)
Applicable modules: U8554, LR8534
Measuring strain
See “11.2 Measuring Strain” (p. 270).
Converting measured values
using scaling based on the
gage factor or a strain gage-
type converter’s rated value
See “1.5 Using the Scaling Function” (p. 42).
Connecting a strain gage or
strain gage-type converter
See “Connecting a strain gage or strain gage-type converter” in the
Quick Start Manual.
>
>
3
3
1
1
4
4
5
5
2
2
1
Select the module (Unit) and channel to configure and select the checkbox.
Measurement will not be performed for channels whose checkboxes are not selected.
2
Select the waveform display color.
×
(OFF), 24 colors
Select
[×]
if you wish to measure the channel but not to display its waveform or numerical values on the
screen.
3
Set the input type to
[Strain]
.
4
Under
[Range]
, select the measurement range as appropriate for the measurement target.
1000 µε
,
2000 µε
,
5000 µε
,
10000 µε
,
20000 µε
,
50000 µε
,
100000 µε
,
200000 µε
The instrument expresses strain in terms of micro epsilon (μ
ε
).
1
Settings and Operation
Summary of Contents for LR8450
Page 68: ...64 Observing Waveforms Alarm...
Page 72: ...68 Observing Waveforms 3 Numerical calculation Numerical calculation results...
Page 131: ...128 Acquiring Data with a Computer PC...
Page 175: ...172 Controlling the System...
Page 233: ...230 Controlling the Instrument with Communication Commands...
Page 271: ...268 Plug in Module Specifications...