141
Assigning Event Marks with an External Signal
Unit
Measure
Channel
Trigger
Alarm
Calculation
System
5.2 Assigning Event Marks with an External Signal
You can assign event marks by inputting external signals.
This function must be configured before measurement begins.
>
>
2
2
1
1
1
Under
[External input 1]
to
[External input 3]
, select
[Event input]
.
2
Under
[Start]
, select the slope at which to assign the event mark.
↑
Assigns the event mark at the pulse’s rising edge.
↓
Assigns the event mark at the pulse’s falling edge.
3
Input the pulse to the input terminal (I/O 1 to I/O 3) with the
[Event input]
setting.
An event mark and number will be assigned at the edge of the entered pulse.
High
(2.5 V to 5.0)
Low
(0 V to 0.8 V)
See “External control connections” in the Quick Start Manual and “8 External Control (EXT. I/O)”
5
Marking Functionality