107
Saving Data
Unit
Measure
Channel
Trigger
Alarm
Calculation
System
>
>
2
2
3
3
1
1
4
4
5
5
1
Enter the filename to use during auto-save operation in the
[File name]
field (up to 8 single-
byte characters).
See “Text entry method” (p. 8).
A serial number beginning from 0001 (in increments of one) will be appended to entered file names.
Example: Filename:
[ABC]
, Format :binary
ABC0001.MEM, ABC0002.MEM, ABC0003.MEM, . . .
Example: Filename:
[ABC100]
, Format: binary
ABC1000001.MEM, ABC1000002.MEM, ABC1000003.MEM, . . .
If you do not specify a filename, a filename will be assigned automatically.
See “When saving files without specifying a filename” (p. 100).
Under
[Add trigger date/time],
select whether to add date and time to the filename.
Adds the date and time at which the trigger activated to the waveform data filename.
Does not add the trigger activation date and time the end of the waveform data
filename.
When the
[Add trigger date/time]
check box is selected, the file is named as below.
Entered filename
_191224_235959_0001.MEM (when the trigger activated at 23:59:59 on December 24, 2019)
2
Under
[Media]
, select the media to give priority when saving data.
SD card
,
USB flash drive
When both an SD Memory Card and USB Drive are inserted, the system will save data on the selected media.
If the selected media is not inserted, the system will save data on the other media.
3
Saving and Loading Data
Summary of Contents for LR8450
Page 68: ...64 Observing Waveforms Alarm...
Page 72: ...68 Observing Waveforms 3 Numerical calculation Numerical calculation results...
Page 131: ...128 Acquiring Data with a Computer PC...
Page 175: ...172 Controlling the System...
Page 233: ...230 Controlling the Instrument with Communication Commands...
Page 271: ...268 Plug in Module Specifications...