24
Configuring Input Channels
Measuring temperature (with resistance temperature detectors)
This section describes how to configure settings on the individual settings screen when measuring
temperature using resistance temperature detectors.
You can use
[Input]
on the settings list screen to configure the settings. (See p. 50.)
Applicable modules: U8551, LR8531
>
>
3
3
5
5
1
1
4
4
6
6
2
2
1
Select the module (Unit) and channel to configure and select the checkbox.
Measurement will not be performed for channels whose checkboxes are not selected.
2
Select the waveform display color.
×
(OFF), 24 colors
Select
[×]
if you wish to measure the channel but not to display its waveform or numerical values on the
screen.
3
Set the input type to
[RTD]
.
4
Under
[Type]
, select the type of resistance temperature detector you’re using.
Pt100
,
JPt100
,
Pt1000
When
[Pt1000]
is selected, the
[10 ms]
,
[20 ms]
, and
[50 ms]
data refresh interval settings will not be
available.
5
Under
[Range]
, select the measurement range as appropriate for the temperature being
measured.
100°C
,
500°C
,
2000°C
6
Under
[Wiring method]
, select a the wiring method of the resistance temperature detector.
3-wire
3-wire resistance temperature detector.
4-wire
4-wire resistance temperature detector.
Summary of Contents for LR8450
Page 68: ...64 Observing Waveforms Alarm...
Page 72: ...68 Observing Waveforms 3 Numerical calculation Numerical calculation results...
Page 131: ...128 Acquiring Data with a Computer PC...
Page 175: ...172 Controlling the System...
Page 233: ...230 Controlling the Instrument with Communication Commands...
Page 271: ...268 Plug in Module Specifications...