137
Checking Alarms
Unit
Measure
Channel
Trigger
Alarm
Calculation
System
4.2
Checking Alarms
This section describes how to check whether any alarms have occurred on the
[Alarm]
screen.
1
Press the
WAVE
key several times to display the
[Alarm]
screen.
22 33
44
55
66
77
2
2
4
4
1
1
3
3
1
1
ALM1
to
ALM8
Red: Alarm output; Green: No alarm output
22
No.
Alarm memory number (assigned in order that alarms occur, starting with 1)
33
ALM
Alarm number (ALM1 to ALM8)
44
UNIT-CH
Module and channel number
55
Error
Type of error (thermocouple wire break, wireless communications error, low
battery)
66
Time of occurrence
Time at which the alarm occurred
77
Time of cancellation
Time at which the alarm was cleared
•
The alarm memory stores up to 100 alarms.
• The communication interruption log (event record) is maintained regardless of alarm settings.
• Alarms occurring after the start of measurement are recorded in the log.
2
Under
[Display alarm]
, select the alarm number for which you wish to display an alarm
history.
The history for the selected alarm number (
[ALM1]
to
[ALM8]
) will be displayed.
ALL
,
ALM1
,
ALM2
,
ALM3
,
ALM4
,
ALM5
,
ALM6
,
ALM7
,
ALM8
To display all alarms, select
[ALL]
.
Scroll through the alarm history using the
SCROLL/CURSOR
keys.
3
(When you wish to review the waveform from when the alarm occurred)
Specify the alarm memory number and press the
ENTER
key while
[Jump]
is selected.
The waveform starting at the time the specified alarm occurred will be displayed.
4
Alarm (Alarm Output)
Summary of Contents for LR8450
Page 68: ...64 Observing Waveforms Alarm...
Page 72: ...68 Observing Waveforms 3 Numerical calculation Numerical calculation results...
Page 131: ...128 Acquiring Data with a Computer PC...
Page 175: ...172 Controlling the System...
Page 233: ...230 Controlling the Instrument with Communication Commands...
Page 271: ...268 Plug in Module Specifications...