Scan Timing
U8550, U8551, LR8530, and LR8531
The following diagrams provide examples of typical scan timing for the U8550, U8551, LR8530,
and LR8531.
Example: Data refresh interval of 10 ms, measurement on for all 15 channels, wire
break detection off
Channels 1 to 15 are scanned with a duration of approximately 0.5 ms per channel.
The scan starts again with CH1 after the data refresh interval of 10 ms has elapsed.
Data refresh interval: 10 ms
About 0.5 ms
CH1
CH2
CH3
CH13
CH14
CH1
CH2
CH15
Example: Data refresh interval of 10 ms; measurement on for CH1, CH5, and
CH15; wire break detection off
Only channels for which measurement is enabled are scanned; channels for which measurement is
disabled are not scanned.
Data refresh interval: 10 ms
About 0.5 ms
CH1
CH5
CH15
CH1
CH5
Example: Data refresh interval of 5 s, measurement on for all 15 channels, wire
break detection off
Channels 1 to 15 are scanned with a duration of approximately 160 ms or 180 ms per channel,
depending on the power supply frequency filter setting.
Data refresh interval: 5 s
About 180 ms (power supply frequency filter: 50 Hz)
CH1
CH2
CH3
CH13
CH14
CH1
CH2
CH15
About 160 ms (power supply frequency filter: 60 Hz)
11
Knowledge and Information
Summary of Contents for LR8450
Page 68: ...64 Observing Waveforms Alarm...
Page 72: ...68 Observing Waveforms 3 Numerical calculation Numerical calculation results...
Page 131: ...128 Acquiring Data with a Computer PC...
Page 175: ...172 Controlling the System...
Page 233: ...230 Controlling the Instrument with Communication Commands...
Page 271: ...268 Plug in Module Specifications...