35
Configuring Input Channels
Unit
Measure
Channel
Trigger
Alarm
Calculation
System
Treatment of data that exceeds the measurable range
Regardless of the measurement target, measured values that exceed the measurable range are
treated as over-range values, resulting in a display of either
[+OVER]
or
[-OVER]
on the numerical
value display and for A/B cursor values.
Treatment of such values in saved data and calculation results is described in “11.13 Data
The screen will display a waveform saturated by the value described in “11.13 Data Handling”
1000
400
0
- 200
- 1000
(°C)
When using wire break detection in thermocouple measurement, values are treated as follows:
• When there is a wire break or the thermocouple’s measurable range is exceeded in the positive
direction, the value is treated as a wire break (BURNOUT).
• When the thermocouple’s measurable range is exceeded in the negative direction, the value is
treated as (-OVER).
1
Settings and Operation
Summary of Contents for LR8450
Page 68: ...64 Observing Waveforms Alarm...
Page 72: ...68 Observing Waveforms 3 Numerical calculation Numerical calculation results...
Page 131: ...128 Acquiring Data with a Computer PC...
Page 175: ...172 Controlling the System...
Page 233: ...230 Controlling the Instrument with Communication Commands...
Page 271: ...268 Plug in Module Specifications...