Using the Scaling Function
1.5 Using the Scaling Function
This section describes how to use the scaling function to convert voltage values measured by the
instrument into the measurement target’s physical properties (current, temperature, etc.) and then
display or record them.
Converted values can be displayed using decimal or scientific notation.
Before scaling
After scaling
[V]
[A]
0.2
0.1
0
10
5.0
0
>
>
4
4
3
3
1
1
2
2
1
Select the scaling display method.
OFF
Does not use the scaling function.
Decimal
Displays converted values as decimals.
Exponent
Displays converted values as exponents.
2
Select the scaling conversion method.
Ratio
Sets the physical quantity (conversion ratio) per 1 V of input signal and the offset.
2-point
Sets converted values for two input signal voltage values.
Sensitivity
Sets the sensitivity constant for a heat flow sensor or actinometer.
Rating
Sets the rated capacity and rated output according to values in the inspection results
sheet for the strain gage-type converter.
(Setting available when using the U8554/LR8534 Strain Unit)
The conversion method cannot be selected for pulse integration measurement.
See “Scaling settings during integration measurement” (p. 45).
Summary of Contents for LR8450
Page 68: ...64 Observing Waveforms Alarm...
Page 72: ...68 Observing Waveforms 3 Numerical calculation Numerical calculation results...
Page 131: ...128 Acquiring Data with a Computer PC...
Page 175: ...172 Controlling the System...
Page 233: ...230 Controlling the Instrument with Communication Commands...
Page 271: ...268 Plug in Module Specifications...