background image

 

 

SciLog

®

 DINCon

®

, DINPres

®

& DINTemp

®

 Modules 

Installation, Operating & Maintenance Instructions

Summary of Contents for SciLog DINCon

Page 1: ...SciLog DINCon DINPres DINTemp Modules Installation Operating Maintenance Instructions...

Page 2: ...erations This publication is for general information only and customers are requested to contact your local Parker Sales Representative for detailed information and advice on a product s suitability f...

Page 3: ...7 Maintenance Cleaning 7 DIN Module Specifications 8 Sensor Specifications 9 SciCon Sensors 9 SciPres Sensors 9 SciTemp Sensors 10 Installation 11 Operating Instructions 14 Modbus Register List 16 DI...

Page 4: ...WER TO THIS UNIT BEFORE INSTALLING SERVICING CLEANING OR REMOVING THE FUSE FAILURE TO DO SO COULD RESULT IN BODILY HARM AND OR PROPERTY DAMAGE CAUTION OBSERVE PRECAUTIONS FOR HANDLING ELECTROSTATIC SE...

Page 5: ...TRETIEN LE NETTOYAGE OU AVANT DE RETIRER LE FUSIBLE NE PAS OBSERVER CES PR CAUTIONS RISQUERAIT DE CAUSER DES BLESSURES CORPORELLES OU ET D ENDOMMAGER L QUIPEMENT PRUDENCE SOYEZ PRUDENT LORSQUE VOUS MA...

Page 6: ...tration 2340 Eastman Avenue Oxnard California USA 93030 toll free 877 784 2234 phone 1 805 604 3400 fax 1 805 604 3401 email dhpsales na parker com www parker com dhsingleuse Parker Hannifin Manufactu...

Page 7: ...chnical specifications and input ratings of the SciLog sensors are observed See the appropriate Sensor Specifications page The protection provided by this equipment may be impaired if the SciLog senso...

Page 8: ...n 3 N C 4 Signal GND COM 5 RS 232 RX 6 RS 232 TX 7 RS 485 A 8 RS 485 B I O Port 2 RJ45 Ethernet Connection I O Port 3 USB Connection Software I O Port Bottom 8 Pin Sensor Connection DINCon has 1 DINPr...

Page 9: ...be Thermistor Pre Calibration 0 10 M KCI Solution 12 88 mS at 25 0 C Sensor Microchip EPROM Stores Sensor ID and Calibration Factor Sensor Connector Cables Lockable and Waterproof SciPres Power 5 VDC...

Page 10: ...ded by DINTemp Module Fluid Connections Choice of five sizes Luer 3 8 Barb 1 2 Barb 3 4 Tri Clover TC Sanitary and 1 Tri Clover Ladish Sanitary Wetted Materials Polysulphone meets USP Class VI and FDA...

Page 11: ...the bottom Wiring Power and Communication Connect 24 Volts DC 250 mA power to Pins 1 and 2 of the 8 pin terminal block Note pin 1 is at the rear All data acquisition is via Modbus registers in either...

Page 12: ...ctor on the sensor end and an 8 pin connector on the other end that plugs into a matching port on the bottom of the Module The DINCon has only one port and uses only one SciCon Sensor The DINPres and...

Page 13: ...as follows Green Connected Red Disconnected Yellow Connected but not calibrated If the LED shows Yellow it s time to replace the affected Sensor with a new one Connecting the wrong type of Sensor to t...

Page 14: ...dule communications RS232 Modbus RTU or Modbus ASCII RS485 Modbus RTU or Modbus ASCII Ethernet Modbus TCP USB Modbus RTU or Modbus ASCII The active Modbus interface is selected by connecting to the mo...

Page 15: ...he ID numbers can be found in registers 201 206 for Sensor 1 241 246 for Sensor 2 and 281 286 for Sensor 3 Note The first letter of this ID number a C S or T is not in any of the registers Units and A...

Page 16: ...he DINPres reads and utilizes calibration data from SciPres Sensors that have been calibrated to two points 0 and 30 psi or three points 0 54 and 6 psi depending upon the type of stock purchased It ha...

Page 17: ...xa55a hex 0xa55a registers 104 109 must be written as a block 105 contains Modbus address 10 106 R W 16 bit Modbus Mode 0 0 RS485 RTU 1 RS485 ASCII 2 RS232 RTU 3 RS232 ASCII 4 USB RTU 5 USB ASCII 6 TC...

Page 18: ...35 R W 16 bit Hour 24 136 R W 16 bit Minute 137 R W 16 bit Second 140 R W 16 bit Conductivity Units 0 0 uS 1 mS 2 ppm KCl 3 ppm NaCl 4 ppm 442 141 142 R W IEEE Float Cal Concentration default 12 88mS...

Page 19: ...it Temperature Comp Disabled if NZ 0 0 201 R W 16 bit Sensor ID Type 0 Luer 1 3 8 Barb 2 1 2 Barb 3 3 4 TC 4 1 TC 100 Cal Fixture 202 R W 16 bit Sensor ID Material 0 Polypropylene 1 Polysulfone 203 R...

Page 20: ...9 must be written as a block 105 contains Modbus address 20 106 R W 16 bit Modbus Mode 0 0 RS485 RTU 1 RS485 ASCII 2 RS232 RTU 3 RS232 ASCII 4 USB RTU 5 USB ASCII 6 TCP 107 R W 16 bit Modbus Baud Rate...

Page 21: ...W 16 bit Month 133 R W 16 bit Day 134 R W 16 bit Year 135 R W 16 bit Hour 24 136 R W 16 bit Minute 137 R W 16 bit Second 140 R W 16 bit Pressure Units 0 0 psi 1 kPa 2 bar 3 mmHg 141 142 R W IEEE Floa...

Page 22: ...Limit High 0 0 168 169 R W IEEE Float P3 Pressure Alarm Limit Low 0 0 170 171 R W IEEE Float P3 Pressure Alarm Limit High 0 0 201 R W 16 bit P1 Sensor ID Type 0 Luer 1 3 8 Barb 2 1 2 Barb 3 3 4 TC 4...

Page 23: ...D Year 245 R W 16 bit P2 Sensor ID Lot 246 R W 16 bit P2 Sensor ID Unit Number 248 R W IEEE Float P2 Sensor Pressure Offset PZ 250 R W IEEE Float P2 Sensor Cal Factor CF 252 R W IEEE Float P2 Sensor C...

Page 24: ...al Factor CF 292 R W IEEE Float P3 Sensor Cal Pressure CP1 294 R W IEEE Float P3 Sensor Cal Factor 2 CF2 296 R W IEEE Float P3 Sensor Cal Pressure 2 CP2 300 R W 16 bit P3 Sensor Custom Cal Status bit...

Page 25: ...9 must be written as a block 105 contains Modbus address 30 106 R W 16 bit Modbus Mode 0 0 RS485 RTU 1 RS485 ASCII 2 RS232 RTU 3 RS232 ASCII 4 USB RTU 5 USB ASCII 6 TCP 107 R W 16 bit Modbus Baud Rate...

Page 26: ...136 R W 16 bit Minute 137 R W 16 bit Second 140 R W 16 bit Temperature Units 0 0 C 1 F 141 142 R W IEEE Float Cal Temperature Default 25C Only change for use with a Custom Calibration 25 0 150 R 16 b...

Page 27: ...C 100 Cal Fixture 202 R W 16 bit T1 Sensor ID Material 0 Polypropylene 1 Polysulfone 203 R W 16 bit T1 Sensor ID Month 204 R W 16 bit T1 Sensor ID Year 205 R W 16 bit T1 Sensor ID Lot 206 R W 16 bit T...

Page 28: ...D Type 0 Luer 1 3 8 Barb 2 1 2 Barb 3 3 4 TC 4 1 TC 100 Cal Fixture 282 R W 16 bit T3 Sensor ID Material 0 Polypropylene 1 Polysulfone 283 R W 16 bit T3 Sensor ID Month 284 R W 16 bit T3 Sensor ID Yea...

Page 29: ...w Gold Electrodes Medical Grade meets USP Class VI and FDA 21 CFR 177 1520 All Wetted Materials are made of Animal Free Compounds Most Sanitizing Agents such as NaOH Hypochlorite Sensors can be autocl...

Page 30: ...Sensor Type Medical grade silicone piezoresistive sensing element with on chip temperature compensation Sensor Isolation Insoluble silicone dielectric gel isolates sensing element from process solutio...

Page 31: ...and FDA 21CFR177 1520 All Wetted Materials are made of Animal Free Compounds Compatible with Most Sanitizing Agents such as NaOH Hypochlorite Sensors can be autoclaved Sensor Type Thermistor Epoxy Co...

Page 32: ...ls For this test SciCon sensors were randomly selected from Parker inventory Group 1 SciCon Sensor Conductivity Response after Exposure to 1 0 N NaOH Solution Temperature 1 0 N NaOH at 22 C 2 Hours Gr...

Page 33: ...ibrated SciCon conductivity sensors 5 were removed from inventory and tested against a conductivity solution 12 99 mS in a temperature equilibrated glove box The out of box sensor response data is lis...

Page 34: ...ed with any of the listed sanitizing solutions Do not exceed the recommended number of of sanitization cycles Sensor accuracy and precision will be affected adversely by excessive cleaning sanitizatio...

Page 35: ...t as well as the 8 pin sensor connector were covered with autoclavable tape Cole Parmer P N EG 08277 62 The sensors were placed into a paper bag 6 1 2 x 4 x12 3 8 sealed with tape and placed in a Tutt...

Page 36: ...distilled water 2 SciTemp sensors can be repeatedly sanitized with any of the listed sanitizing solutions Do not exceed the recommended number of sanitization cycles Sensor accuracy and precision will...

Page 37: ...e The SciTemp sensor air vent as well as the 8 pin sensor connector were sealed with autoclavable tape Cole Parmer P N EG 08277 62 The sensors were placed into a paper bag 6 1 2 x 4 x12 3 8 sealed wit...

Page 38: ...based on testing from ETL C2 11 19 09 A Dawson Addition of note in Modbus section and correction of register 109 settings to reflect pre set stop bit values even odd 1 none 2 C3 3 26 10 A Dawson Corre...

Page 39: ...39 Notes...

Page 40: ...40 Notes...

Page 41: ...41 Notes...

Page 42: ...Durham DH3 2SF England phone 44 0 191 4105121 fax 44 0 191 4105312 email bioscience emea parker com www parker com bioprocessing Parker Hannifin Corporation Bioscience Division N A 2340 Eastman Avenu...