198
_____________________________________________________________________________________________
8.4 Input Terminals
______________________________________________________________________________________________
8.4.3 External Sampling <EXT_SAMP Terminal (A8)>
More than 500
μ
s
More than 500
μ
s
More than 1 ms
E X T _ S A M P
500μs以上
500μs以上
1ms以上
More than 500
μ
s
More than 500
μ
s
More than 1 ms
A trigger signal may be input from an external device.
The EXT_SAMP terminal (terminal No.: A8) initiates sampling with the
rising of a signal (active HIGH).
The maximum delay from the falling edge until activation is 100
μ
s.
Input Signal
Input a signal rising from LOW to HIGH to EXT_SAMP terminal (A8).
Sampling is done at a rising edge.
Input signals of up to 1KHz are available to activate the trigger.
<When using an internal power supply>
HIGH level voltage: Max.: 5.0 [V], Min.: 4.0 [V]
LOW level voltage : Max.: 1.0 [V], Min: 0 [V]
<When using an external power supply>
External power supply voltage V
EXT
: Max.: 30 [V] DC, Min.: 5 [V] DC
HIGH level voltage : Max.: V
EXT
[V], Min.: V
EXT
-1.0 [V]
LOW level voltage : Max.:
1.0 [V],
Min.: 0 [V]
<Timing>