133
_____________________________________________________________________________________________
5.5 Value Comparison Details
______________________________________________________________________________________________
0
t
指定レベルL
T
Specified
level
0
1
2
3
指定レベル
Specified
level
最大値
最小値
Max.
Min.
NOTE
The calculation results of (9) to (12) above may not be displayed, depending on
the waveform.
When scaling is ON, calculations are made after the waveform data is scaled.
The unit will be the one set for scaling.
(19) Level Time (Specified Level Time)
Calculates time [in seconds] from a trigger event to the waveform first
crossing the specified level in the selected direction.
(If the wave crosses the specified level before the trigger due to pre-
trigger, this result will be a negative value.)
(Specified level) = L (Specified time) =
t
The diagram below shows an up-slope case.
(20) Pulse Count
Counts the number of times the waveform crosses the specified level in
the selected direction and then crosses again in the opposite direction.
The diagram below shows an up-slope case. When UP is selected, the
width of the "H" pulse is calculated, and when DOWN is selected, the
width of the "L" pulse is calculated.
(21) MAX & MIN
Obtains the maximum and minimum values of the stored waveform.