192
_____________________________________________________________________________________________
8.2 Terminals
______________________________________________________________________________________________
DANGER
To avoid damage to the instrument, do not apply voltage exceeding
the followings.
Maximum Input Voltage: 30 V DC.
Maximum Voltage to earth: 30 V rms or 60 V DC.
External Control Terminals
Nameplate
8.2 Terminals
The 8730-10, 8731-10 has 32 external control terminals. Input terminals:
12; output terminals: 16; power supply/GND: 4
The external control terminals are isolated from the 8730-10, 8731-10 by a
photocoupler.
M3 screws are used to secure the terminals.
Lift the terminal block cover to expose the terminals.
The cover contains a terminal nameplate. The rear of this nameplate is
left blank so that you can fill in names or functions there.
Terminal name labels are provided with the 8730-10, 8731-10. Attach the
labels to the 8730-10, 8731-10, as required. Never attach labels on the
ventilation slot on the side of the 8730-10, 8731-10.